Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add filters








Language
Year range
1.
Yonsei Medical Journal ; : 1104-1109, 2013.
Article in English | WPRIM | ID: wpr-198367

ABSTRACT

PURPOSE: For the successful completion of transcatheter closure of atrial septal defects with the Amplatzer septal occluder, shape of the defects should be considered prior to selecting the device. The purpose of this study is to evaluate the results of a transcatheter closure of oval shaped atrial septal defect. MATERIALS AND METHODS: From November 2009 until November 2011, cardiac computed tomography was performed on 69 patients who needed a transcatheter closure of atrial septal defect. We defined an oval shaped atrial septal defect as the ratio of the shortest diameter to the longest diameter 0.05). The differences between the device size to the longest diameter of the defect and the ratios of the device size to the longest diameter were significantly smaller in the ovoid group (1.8+/-2.8 vs. 3.7+/-2.6 and 1.1+/-0.1 vs. 1.2+/-0.2). CONCLUSION: Transcatheter closure of an oval shaped atrial septal defect was found to be safe with the smaller Amplatzer septal occluder device when compared with circular atrial septal defects.


Subject(s)
Humans , Middle Aged , Echocardiography , Heart Septal Defects, Atrial/pathology , Postoperative Complications , Republic of Korea , Retrospective Studies , Septal Occluder Device/adverse effects , Tomography, X-Ray Computed , Treatment Outcome
SELECTION OF CITATIONS
SEARCH DETAIL